Home RESEARCH Equipment

LOI (Low Oxygen Index) Chamber

 

Name of equipment:

LOI (Low Oxygen Index) Chamber

Purpose:

Instrument for LOI determination of plastics, polymers and fabrics.

Model, type, country of origin:

Dynisco, USA

Technical characte-ristics:

Weight: 14,7 kg

Dimension: 50,8 x 30,5 x 31,5 cm

Year of origin:  2005                   

Figure:

 

Method of
work:

LOI chamber accurately determines the relative flammability of plastics and other materials by measuring the minimum oxygen concentration that will support combustion. The testing specimens are burned in a precisely controlled atmosphere of nitrogen and oxygen. The operator adjusts the supply gases and uses the flowmeter readings to calculate the oxygen index.

Department,
room, location:

Department for Textile Chemistry and Ecology, PhD Research Cabinet, Savska cesta 16/9

Addition note:   /

 
r1
TTF
r3
T-POT
r5
FP-7
r3
STEP
r8